SV 971A User Manual Appendixes
202
(1) flag record
< flags > :
b0: 1- the overload detected, 0 - the overload not detected
b1: 1- the excessive self-vibration detected, 0 - the excessive self-vibration overload not detected
(2) results of the measurement from the first profile if the corresponding LOGGER position was active
(paths: Measurement / Logging / Logger Res. / Prof. 1); up to seven words are written:
<result1> - Lxpeak
1
result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result2> - Lxymax
2
result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result3> - Lxymin
2
result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result4> - Lxyeq
23
result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result5> - LAV result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result6> - LR1 result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
<result7> - LR2 result, depending on the value of BufferP[1] (cf. Tab. B.1.10)
(3) results of the measurement from the second profile if the corresponding LOGGER position was
active (paths: Measurement / Logging / Logger Res. / Prof. 2); up to five words are written:
<result1> - Lxpeak
1
result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result2> - Lxymax
2
result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result3> - Lxymin
2
result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result4> - Lxyeq
23
result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result5> - LAV result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result6> - LR1 result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
<result7> - LR2 result, depending on the value of BufferP[2] (cf. Tab. B.1.10)
(4) results of the measurement from the third profile if the corresponding LOGGER position was active
(paths: Measurement / Logging / Logger Res. / Prof. 3); up to five words are written:
<result1> - Lxpeak
1
result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result2> - Lxymax
2
result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result3> - Lxymin
2
result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result4> - Lxyeq
23
result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result5> - LAV result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result6> - LR1 result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
<result7> - LR2 result, depending on the value of BufferP[3] (cf. Tab. B.1.10)
1
x - depends of the filter type for Peak result calculation in selected profile: A, C, Z, B (cf.
Tab. B.1.10)
2
x - depends of the filter type in selected profile: A, C, Z, B (cf. Tab. B.1.10)
y - depends of the detector type in selected profile: I (imp.), F (fast), S (slow) (cf. Tab. B.1.10)
3
y - only for exponential detector's type (cf. Tab. B.1.6)
(5) results of 1/1 Octave analysis or 1/3 Octave analysis if 1/1 Octave analysis or 1/3 Octave analysis
was selected as the measurement function and the Logger was active (paths: Measurement /
Logging / Logger Res. / Peak Sp. [√] and Leq Sp. [√] ); the sequence of words is written:
<Octave Peak[1]> <Octave Peak [2]> ... <Octave Peak [Noct+NOctTot]> <Octave Leq[1]> <Octave
Leq[2]> ... <Octave Leq[NOct+NOctTot]>
where:
Octave Peak[i] - the result of 1/1 Octave or 1/3 Octave Peak analysis (*100 dB);
i = 1..NOct+NOctTot
Octave Leq[i] - the result of 1/1 Octave or 1/3 Octave Leq analysis (*100 dB);
i = 1..NOct+NOctTot