2212
Service Manual
Performance Tests
Procedure
Steps
Step 1.Check Probe Adjust Operation
a.
SET:
VOLTS/DIV
(CH
1)
........ 10 mV
SEC/DIV
...
.. ..
..
.....
.. .. ..
...
..
0.
5 ms
Trigger SOURCE............
CH
1
b.
Connect the 10X probe to
the
CH
1
OR
X
input
connector and
clip the probe
tip
to the
PROBE
ADJUST
connector on the
front panel.
c.
If necessary, adjust the probe compensation for a flat-topped
squarewave display.
d.
CHECK
-That
the display amplitude is between 4.75 to 5.25
divisions.
4-
23