Adjustment Procedures
Trigger Adjusbnents
5-
16
Equipment Required
SG502 (item 4 )
PG506A
DC
generator (item 17)
BNC
cables (item
7)
Initial Control Settings
Calibration
MODE
Adjust Trigger
Procedure
Calibration
via
the
Calibration Routine
in
the
2212
a.
Place:
CAL
jumper in
CAL
position, and
wait
for
the
Calibration
MENU
on the
CRT
screen.
b.
Select:
TRIGGER
to
GO.
c.
Adjust: Trigger by following the instructions at the screen.
d.
Remove:
CAL
jumper to enter normal scope operation.