Adjusbnent Procedures
Position Adjushnents Equipment Required
TG501 (Item 3)
Cable 50 n (Item
7)
2212 Service Manual
50 .Q
BNC
Termination (Item 11)
Initial Control Settings
Calibration
MODE
Adjust Position
Procedure
Calibration
via
the Calibration Routine
in
the 2212
a.
Place:
CAL
jumper
in
CAL
position,
and
wait
for
the
Calibration MENU on the
CRT
screen.
b. Select: POSITION to
GO.
c.
Adjust: POSITION by following the instructions at the screen.
d. Remove:
CAL
jumper to enter normal scope operation.
5-
21