Measurement setups Measurement Setup control window (Time tab)
Positive Overshoot
This voltage measurement is measured over the entire waveform o r gated region.
Positive Overshoot = (Maximum High) × Amplitude 100%
Negative Oversho ot
This voltage measurement is measured over the entire waveform or ga t ed region and is expr essed as:
Negative O vershoot = (Low Minimum) × Amplitude × 100%
Mean
This voltage measurement is the arithmetic mean over the entire waveform o r gated region.
Cycle Mean
This voltage measurement is the arithmetic mean over the rst cycle in the waveform or the rst
cycle in the gated region.
Measurement Setup control w indow (Time tab)
From the Measure menu, select Measurement Setup; then open the Time tab.
Overview
Use this control window to set up the instrument for taking automatic time measurements.
To use
For information o n the controls, click the buttons.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 273

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals