Vertical Setups Enhanced Bandwidth
Overview
The Bandwidth Enhancement controls (similar to those shown below) allow you to use DSP processing
to achieve more accurate rise time measurements, to extend the bandwidth, and to atten the passband
at the full sample rate.
Enhanced Bandwidth also provides a matched response across enabled channels for channel-to-channel
comparisons and differential measurements. The available bandwidth enhancement s elections will vary
depending on the instrument you purchased. Enhanced bandwidth is extended to the probe tip for some
high performance probes and tips.
To use
To enable Bandwidth Enhancement for a vertical input channel:
1. Select a Channel tab in the Vertical Setup window.
2. Click Digital Filters (DSP) Enabled to activate the Bandwidth Enhancement lters.
3. Click Apply to All Channels to On to apply Bandwidth Enhancement to all channels simultaneously.
4. Click Analog Only to disable Bandwidth Enhancement and return to Analog operation. The Bandwidth
drop-down list adds (HW) to the selection indicating a hardware/analog-only lter.
The following screen appears on MSO/DPO5000 Series instruments:
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 495

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals