Oscilloscope Reference Measurement algorithms
Details of the integration algorithm are given later. Integration Algorithm (see page 737)
Burst width
Timing measurement. The duration of a burst.
1. Find MCross1 on the waveform. This is MCrossStart.
2. Find the last MCross (begin the search at EndCycle and search toward StartCycle). This is MCrossStop.
This could be a different value from MCross1.
3. Compute BurstWidth = MCrossStop MCrossStart
Cycle m ean
Amplitude (voltage) measurement. The mean over one waveform cycle. For non-cyclical data, you might
prefer to use the Mean measurement.
If StartCycle = EndCycle then return the (interpolated) value at StartCycle.
Details of the integration algorithm are given later. Integration Algorithm (see page 737)
Cycle RMS
The true Root Mean Square volt
age over one cycle.
If StartCycle = EndCycle then CycleRMS = Waveform[Start].
Otherwise,
Details of the integration algorithm are given later. Integration Algorithm (see page 737)
730 DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals