Reference Error Codes and Warnings
Error Codes
Error Message Description
Possible Solut
ions
Utilities
E701
Number of edges found in
the waveform i
s less than
required edges (user input)
to perform the deskew
operation
Before starting the deskew
operation, th
e application
tries to nd out if the
edge(s) available for
deskew opera
tion is greater
than or equal to the number
of edges needed
Adjust the timebase and
(or) increase the record
length. Perfo
rm the deskew
operation by entering less
number of edges
E702 Propagation delay is more
than one cycle
The deskew operation is
performed on the same
signal sour
ce. Check
whether dissimilar signals
or any external sources
(like diff
erent probes) are
used
Check whether the same
signal sour
ce is used
for the "From" and "To"
channels. Check whether
the bandwi
dth of the probes
is sufcient for the signal
bandwidth
E701
Number of edges found in
the wavef
orm is less than
required edges (user input)
to perform the deskew
operati
on
Before the deskew
operatio
n can start, the
number of edge(s) available
for deskew operation must
be great
er than or equal
to the number of edges
needed
Adjust the timebase and
(or) increase the record
length. P
erform the deskew
operation by entering less
number of edges
E702 Propagation delay is more
than one cycle
The deskew operation is
performed on the same
signal
source. Check
whether dissimilar signals
or any external sources
(like
different probes) are
used
Check whether the same
signal
source is used
for the "From" and "To"
channels. Check whether
the ba
ndwidth of the probes
is sufcient for the signal
bandwidth
TSV File Generator
E721 Error in generating a .tsv
le
This message is generated
if there is an error while
gene
rating the .tsv le for
the given inputs
Check the correctness of
the i
nput .csv le(s) and the
output .tsv le
E72
2
Timestamp of the two
.csv les did not match.
Con
version failed
The
re will be two separate
.csv les, one for D+ and
another for D-. Each of
the
se les consist of the
time values of the D+ and
D- signal. It is expected
th
at both the les s hould
have D ata at same time
stamp
Tr y generating the .tsv le
with a new set of .csv les
158 TDSUSB2 Universal Serial Bus Measurements Package

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDSUSB2 and is the answer not in the manual?

Tektronix TDSUSB2 Specifications

General IconGeneral
BrandTektronix
ModelTDSUSB2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals