6 Using the product
96 0970 4020 en 01 testo AG
Designation Description
Receiver 1 Telephone number of the employee who
is to receive the alarm message. Enter
with country code, network or area code
and local number: 00491721234567.
To acknowledge an alarm
message, send the received
message with the same text back
to the mobile number of the
Saveris base.
Receiver 2 Checkbox and input field for the
telephone number of the employee who is
to receive the alarm message if
Receiver
1
cannot be reached. Enter with country
code, network or area code and local
number: 00491721234567.
To acknowledge an alarm
message, send the received
message with the same text back
to the mobile number of the
Saveris base.
Forward after ...
min
If
Receiver 1 does not react within this
time span, the message is forwarded to
Receiver 2.
Additional text Additional text that is attached to the
alarm message.
The system alarms are above all used for pointing out
irregularities in the measurement system. If no reaction
follows the system alarms, uninterrupted data recording
is not guaranteed.
If a system alarm is issued via
SMS or E-mail time
delays may occur.
The notification function via
SMS is only available if the
Saveris base is equipped with a GSM module.
3. Perform the required settings.
When exiting the basic settings the settings are saved.