EasyManua.ls Logo

Watts AERCO Edge [ii] - Subsystem Calibration

Watts AERCO Edge [ii]
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Edge [ii] Controller Operation Manual
SECTION 3: CALIBRATION
OMM-139_E 8/25/2020 Technical Support (800) 526-0288 Mon-Fri, 8 am - 5 pm EST Page 71 of 196
7. If the output selected in Step 2, includes feedback, it is displayed in the Feedback
parameter (see the Name parameter in Section 6.5.4: Analog Inputs). If no feedback is
sensed, the reading is ---(blank). The feedback signal mapping is:
Analog Output
Analog Input (Feedback)
Display Tag
Cascade Valve
Cas Vlv Fdbk
Valve Feedback
BLR VS (Variable Speed) Pump
BLR VSP Fdbk
Pump Feedback
DHW VS (Variable Speed) Pump
DHW VSP Fdbk
Pump Feedback
8. Repeat Steps 2 through 7 with the other Analog Outputs in use in the system.
3.4 Subsystem
The Subsystem screen provides access to two system components:
The Air/Fuel Valve
The Spark Monitor
Figure 3.4: Calibration: Subsystem Calibration Menu
1. Go to: Main Menu Calibration Subsystem.
2. Press either Air Fuel Valve or Spark Monitoring.
NOTE:
You can navigate between the Air Fuel Valve and Spark Monitor screens in round-robin fashion
using the Page Left and Page Right icons on both screens.

Table of Contents

Related product manuals