<6. PARAMETER DESCRIPTION>
6-32
IM 01E30D01-01EN
(8) Menu J: Auxiliary Function Setting items
MenuJcontainssettingitemssuchastheow
direction, rate limits, and current output limits.
[J10: 4-20mA Low Cut] Setting of the low-cut
range for current output
This parameter is used to force current output to
0 % (i.e., 4 mA) in the vicinity of 0% output, and
setting for the current (4 to 20 mA) output low cut is
madeusingapercentageofthesmallestowrate
span.
Thelowcutsettingvaluedependsonowrate
span.
●Flow rate span < 1 m/s:
Input value or 3 cm/s, whichever is bigger
●Flow rate span ≥ 1 m/s:
Input value
Toconrmtheowratespanvelocityin“m/s”unit,
read [C44: Velocity Check].
Example1:Situationwheretheowratespanisset
to 2 m/s:
When “10 %” is input in this parameter, low cut
setting value is to be “10 %”.
Setting range: 0 to 20%
4mA
Input
0.5%
0.5%
10%
10%
Setting value
Hysteresis
fixed at 1%
5.6mA(10%)
Example2:Situationwheretheowratespanisset
to 0.8 m/s:
When “3 %” is input in this parameter, low cut
setting value is to be “3 cm/s”.
Setting range: 0~20%
4mA
Input
0.5% 0.5%
3%
3cm/s
3cm/s
Input value:3%(=2.4cm/s)
Setting value
Hysteresis
fixed at 1%
If the multiple ranges or forward/reverse ranges is
used, low cut is carried out at the setting value for
the smallest span.
However, the low cut function will be terminated if
this parameter is set to 0 %.
Theindicationsoftheinstantaneousowrate(%,
Actualinstantaneousowrate,mA,Bargraph)on
the display unit are the same action.
NOTE
The default setting is 3%.
When the low cut is small, an incorrect output
mayoccurattheowrateofzero.Usethe
instrument with the default setting 3%. If the span
issmall,thedampingtimeisshortortheuidis
low conductivity, an incorrect output may occur
easilyattheowrateofzero.
[J11: 4-20mA Low Lmt] Setting of the low limit for
current output
This parameter is used to restrict low current
portions of current (4 to 20mA) output, and it is
initially set to -1.25%. Setting should be performed
when a higher value is required for the lower limit.
Theindicationsoftheinstantaneousowrates(%,
Actualinstantaneousowrate,mA,Bargraph)on
the display unit are the same action.
Example: Situation where low limit is set to 10%
4mA
NOTE
When the low limit works, the display indicates
83:Fix Cur Wng as the warning.
NOTE
The set current output at each alarm function
becomes higher priority than set value at J11
when the alarm occurs.
The priority of this output is described below.
The process alarms and setting alarms are
available when “Yes” were selected at G26, G27,
G28, G29 and/or G31.
System alarm (G20) >
Process alarm (G25) >
Setting alarm (G30) >
Low limit (J11)
Example: “3.8mA” is set for G25
The current output is 3.8mA in spite of
setting any value at J11.