EasyManuals Logo

ADVANTEST R3131 Series User Manual

ADVANTEST R3131 Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
R3131 Spectrum Analyzer Operation Manual
2.3.4 Measuring the VA Ratio
ADVANTEST R3131 Fri 19 Dec 1997 Marker(1)
REF 107.0 dBAV ATT 10dB A_max B_blnk
iodpi (75n) Posi Posi Normal
Marker Marker
MARKER 91.25 MHz -------
—92:SCCIBAy Delta
91.25 MHz Marker
Peak â–¶
Menu
-- Sig Track
ON /OFF
Sound â–¶
Marker
Off
CENTER 91.25 MHz SPAN 20.00 MHz 1/2,morelo.
RBW 300 kHz VBW 100 kHz SWP 50 ms
Figure 2-68 Measuring the Video Carrier Level
Measuring the audio carrier level
11. Press MKR, 9, 5, ., 7, 5 and MHz.
The marker is displayed at 95.75 MHz on the trace.
This marker level is called video carrier level A (dBmV).
ADVANTEST R3131
REF 107.0 dB,V
10cipi_c75o)
MARKER
95.77 MHz
Fri 19 Dec 1997 Marker(1)
ATT 10dB A_max B_blnk
CENTER 91.25 MHz
RBW 300 kHz VBW 100 kHz
Posi Posi Normal
Marker Marker
95.77 MHO - --
80.08 ABAV Delta
Marker
Peak â–¶
Menu
Sig Track
ON /OFF
Sound 11,
" Marker
Off
SPAN 20.00 MHz 1/2,more1
SWP 50 ms
Figure 2-69 Measuring the Audio Carrier Level
12. Calculate the VA ratio using the following formula.
VA ratio (dB) = Video carrier level V (dBi_tV) - Audio carrier level A (dBi.i.V)
2-73

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3131 Series and is the answer not in the manual?

ADVANTEST R3131 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3131 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals