Ready For Exposure Status
Table 9: Exposure ready
Color Description
Green
Exposure ready. Indicates that the selected technique is properly set
and there are no interlock failures or system faults.
Red
Exposure not ready.
Check the message frame for more information. It is not possible to
perform an exposure due to an error.
The status will turn to green when problem is solved.
Blue
Exposure not ready.
No examination defined.
108 | DR 400 | Software Console and Tube Head Display