EasyManua.ls Logo

AGFA DR 400 - Generator Technical Data

AGFA DR 400
265 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Generator Technical Data
Manufac-
turer
Spellman High Voltage Electronics GmbH
Josef-Baumann-Strasse 23
D-44805 Bochum, Germany
Supported
Models
EDITOR HFe
401
EDITOR HFe
501
EDITOR HFe
601
EDITOR HFe
801
Max. Pow-
er
40 kW 50 kW 65 kW 80 kW
Power Out-
put (at
0.1s)
500mA:
80kVp
400mA:
100kVp
320mA:
125kVp
266mA:
150kVp
625mA:
80kVp
500mA:
100kVp
400mA:
125kVp
330mA:
150kVp
800mA:
80kVp
650mA:
100kVp
520mA:
125kVp
430mA:
150kVp
800mA:
80kVp
800mA:
100kVp
640mA:
125kVp
530mA:
150kVp
kV-Range
40-150 kV 40-150 kV 40-150 kV 40-150 kV
mAs-Range 0.5-600 mAs 0.5-600 mAs 0.5-600 mAs 0.5-600 mAs
mA-Range 10-500 mA 10-650 mA 10-800 mA 10-800 mA
ms-Range 1-6300 ms 1-6300 ms 1-6300 ms 1-6300 ms
Power line
400 V Y-
source
400V 3N~ PE (Y) 50/60 Hz
Power line
400/480 V
Delta-
source
400/480V 3~PE (delta without N) 50/60Hz
The power setting is selected during installation and printed on
the type label.
Dimen-
sions
89 cm x 43 cm x 29 cm (WxDxH)
Weight 78 kg (400V)
90 kg (400/480V)
Duty cycle The Generator duty cycle is continuous, but limits should be set
during installation depending on the capacity of the X-ray tube.
DR 400 | Product Information | 231
3231E EN 20210708 1530

Table of Contents

Related product manuals