DR 400 Technical Data
Manufacturer Agfa NV
Septestraat 27
2640 Mortsel, Belgium
Type 5520/XXX
Power line 400 V Y-
source
400V 3N~ PE (Y) 50/60 Hz
Power line 400/480
V Delta-source
400/480V 3~PE (delta without N) 50/60Hz
The power setting is selected during installation and
printed on the type label.
Maximum current
(0.2sec) / Power
400V 480V
40 kW generator 92 A / 62 kVA 79 A / 62 kVA
50 kW generator 113 A / 76 kVA 97 A / 76 kVA
65 kW generator 144 A / 96 kVA 124 A / 96 kVA
80 kW generator 180 A / 120 kVA 154 A / 120 kVA
Stand-by power max. 3.3 A
Table movement (full
load of 320 kg)
max. 7.0 A
Permanent filtration
E7254FX X-ray tube 2.8 mm Al @75kVp
(+ 0.2 mm Al with DAP meter integrated in the colli-
mator)
E7884X and E7252X
X-ray tube
2.9 mm Al @ 75kVp
(+ 0.2 mm Al with DAP meter integrated in the colli-
mator)
E7869X X-ray tube 3.1 mm Al @ 75kVp
(+ 0.2 mm Al with DAP meter integrated in the colli-
mator)
DR 400 | Product Information | 229
3231E EN 20210708 1530