System does not start completely if collimator is in
manual mode
Details System does not start completely if collimator is in
manual mode. An error message is displayed indicating
a problem with the collimator during start up.
Cause The key on collimator was not switched back to auto-
matic mode. The system checks communication with
all components during start up. If collimator is in man-
ual mode no communication with the system is availa-
ble.
Brief Solution Set the key at the backside of the collimator to auto-
matic.
Restart the system on X-ray generator console. A restart
of NX is not needed.
DR 400 | Problem solving | 213
3231E EN 20210708 1530