Error numbers
Error
Action
Display Explanation
1
B
tube kV max Tube voltage too high (>166 kV/132 kV)
2
B
control A max Load current too high (> 250 A)
3
B
tube mA max Tube current too high (> 900 A)
4
B
tube +-kV diff Tube voltage difference between +URist
and- URist>15kV
5
B
tube +-mA diff Tube current difference between +IRist
and -IRist >100mA
6
A
ROM test ROM test checksum error
7
A
RAM test RAM test error
8
B
unknown unknown error
9
B
no Tube kV Tube voltage < 10kV after 1ms or < 50%
after 30ms
10
B
tube kV too high Tube voltage > rated voltage + 25%
11
B
inverter overload Converter overload ( > 150000WS)
12
B
send timeout Serial interface transmit timeout
13
A
E²Prom checksum E2PROM checksum error
DR 400 | X-Ray Generator Mini Console | 197
3231E EN 20210708 1530