Error
Action
Display Explanation
14
B
watchdog Watchdog error
15
B
receive timeout Serial interface receipt timeout
16
A
E²Prom wait timeout E2PROM access timeout
17
B
fillament system Heater fault
18
A
DAP system Area dose measuring system self test error
19
A
filament parameter Deviating heating parameters in E2PROM
20
B
+-15V low +-15V outside tolerance
21
B
+5V low +5V outside tolerance
22
B
key is on Key on control panel has been pressed
during switch-on
23
B
XRAY key is on Exposure or fluoroscopy key has been
pressed during switch-on
24
C
mAs max Current time product in mAs has reached
its limit
25
B
exposure too short Not used
26
B
generator not ready Tube switchover error
27
A
service intervall Service interval for maintenance
198 | DR 400 | X-Ray Generator Mini Console