Installation
Installation and configuration is performed by an Agfa trained and authorized
service engineer. Contact your local support organization for more
information.
On a configuration with multiple DR Detectors of the same type, it is required
to apply labeling to the DR Detector containing a unique nickname for each
DR Detector. The nicknames must be configured on the NX Workstation. The
DR Detector Switch shows which DR Detector is active and shows its status, by
means of the nickname of the DR Detector.
An identical label is attached to the bucky of the X-ray system to identify the
dedicated workspace of each DR Detector.
HF-emission and immunity
The HF-emission and immunity can be influenced by connected data cables
depending on length and the manner of installation.
A specific installation environment may require special measures to put the
system into operation according to the remarks for HF-emission and
immunity.
Related Links
Cables, transducers and accessories on page 262
34 | DR 400 | Introduction to DR 400
3231E EN 20210708 1530