Dose Area Product Meter (VacuTec DAP) Technical Data
Manufacturer VacuTec Messtechnik GmbH
Dornblüthstrasse 13
D-01277 Dresden, Germany
Supported Type
VacuDAP 2004
Dose area product range (1.0...9999999.9) cGy x cm
2
Minimum readout value 0.1 cGy x cm
2
Active area
14.7 cm x 14.7 cm
Dimensions 18.2 cm x 17.7 cm x 1.8 cm (WxDxH)
Weight 270 g
Correction factors for using the DAP meter on high altitude
Environmental conditions Correction factor
75 kPa (ca. 2500 m)
0° Celcius
1.26
75 kPa (ca. 2500 m)
20° Celcius
1.31
70 kPa (ca. 3000 m)
0° Celcius
1.35
70 kPa (ca. 3000 m)
20° Celcius
1.40