X-ray generator messages and warning signals
Acoustic signals
The generator indicates particular states with acoustic signals:
Exposure is terminated: 500 ms tone
Errors: rapid series of tones
Visual signals
The generator indicates particular states with visual signals:
Preparation: flashing of prepare ready indicator (green LED)
X-ray tube is prepared: prepare ready indicator is continuously lit (green
LED)
Exposure: radiation indicator is continuously lit (red LED)
Exposure termination
In normal operation the exposure is terminated by the generator when:
mAs product is reached
Exposure time is reached
AEC switches off
If the exposure switch is released the exposure is terminated instantaneously
and an error is indicated.
In case of failure the exposure is instantaneously terminated when:
AEC faulty
Initial dose too high or too low with AEC (if function is activated)
Maximum exposure time of 3.2 sec reached in 1-point technique with AEC
mAs product of 600 mAs is reached
Maximum permissible exposure time of 6.3 sec is reached (safety switch
off)
Door contact is opened