Emergency shutdown power switch
Use the emergency shutdown power switch, if a dangerous situation cannot be
eliminated by pressing the emergency stop button.
WARNING:
Use the emergency shutdown power switch in case of danger to
patients, operators, third parties, or one of the units. The entire
system will be shut down and the power supply will be
disconnected.
The emergency shutdown power switch for the room is typically located on
the wall and easy to access, often close to the power off switch of the X-ray
system. It is installed and labeled by customer.
WARNING:
It must be ensured that the emergency switches are always
freely accessible.
DR 400 | Introduction to DR 400 | 33
3231E EN 20210708 1530