Radiographic Parameters
You can set up following radiographic parameters:
kV: shows the radiographic kV value (X-ray tube voltage) selected for the
exposure.
mAs can show:
The radiographic mAs value selected for the exposure.
When an exposure is made, it shows the actual mAs at the end of the
exposure.
mA: shows the radiographic mA value (current) selected for the exposure.
ms can show:
The time value (in milliseconds) selected for the exposure.
When an exposure is made, it shows the actual time at the end of the
exposure.
Detector ms shows the integration time of the DR detector. When
operating the DR detector, the calculated exposure time (ms) or manual
overrides can never exceed the integration time (detector ms) of the DR
detector.
Max mAs shows the maximum allowed mAs value for exposures using
AEC. The highest allowed setting for max mAs depends on the mA setting
and the detector ms setting. Not available in Free Exposure mode using DR
or Free Exposure mode using CR.
When using AEC, the exposure is terminated by the detector ms or max mAs
settings, even if the target dose is not reached.
118 | DR 400 | Software Console and Tube Head Display
3231E EN 20210708 1530