Collision indicator
Systems with motorized movement have a collision indicator. The collision
indicator avoids collision of the X-ray tube head with the table.
The collision indicator will give a signal in following situations:
• The X-ray tube head is moved manually close than 30 cm to the table top
when performing an examination using the table.
• The X-ray tube head is moved manually closer than 10 cm to the table top
when performing an examination using the wall stand and the X-ray tube
head is rotated toward the wall stand.
The brake is activated and a single beep indicates the collision warning.
To further adjust the position, release the brake button and press it again.
Related Links
X-ray tube stand tracks wall stand height on page 104
DR 400 | Radiographic Table and X-Ray Tube Stand | 135
3231E EN 20210708 1530