Collision protection
Collision protection is only available on the elevating radiographic table.
The collision protection accessories are mounted on the frame of the
radiographic table. They protect the tabletop from damage when colliding
with objects below.
When the collision protection stops downward movement of the radiographic
table, raise the table height and remove the object before lowering the table
again.
Note: The collision protection is influenced by the patient
weight. Take special care when moving the radiographic table
with a patient lying on.
Figure 52: Location of the collision protection accessories
DR 400 | Radiographic Table and X-Ray Tube Stand | 143
3231E EN 20210708 1530