Error
Action
Display Explanation
28
B
no Tube mA Tube current < 50% after 30 ms
29
B/C
tube > 70°C Hood temperature > 70°C
30
-
to save data push 'M' -
31
B
'NOT' signal 'EMERGENCY' safety signal active
32
C
door open Door contact open
33
C
exp. time > 6.3s Exposure time > 6.3 sec
34
B/C
exp. time > 3.2s Exposure time > 3.2 sec (automatic expo-
sure control)
35
B/C
exp. time < 2ms Exposure time < 2 msec (automatic expo-
sure control)
36
C
AEC exposure break Exposure aborted by operator (automatic
exposure control)
37
C
dose too low after
50ms
Dose too small after 50 ms (automatic ex-
posure control)
38
B
pulse delay too long Exposure pause between 2 pulses > 2 sec
(automatic exposure control)
39
C
exp. prepare timeout Exposure preparation timeout
40
B
device ready timeout Device ready timeout
41
B
starter timeout Normal speed starter timeout
DR 400 | X-Ray Generator Mini Console | 199
3231E EN 20210708 1530