Error
Action
Display Explanation
42
B
grid is on Grid active in idle state
43
A
RTC checksum, Batt.
low
RTC (real time clock) error, RTC battery
discharged
44
B
starter system Normal speed starter fault current
45
B
no main current Load current < 4A after 0.5 ms
46
C
exposure stopped by
user
Exposure aborted by operator
47
A
controller - E²prom ver-
ify
Cpu-E2Prom alignment
48
B/C
Wrong tube position Tube position sensor
49
B
Tube mA too high Tube current outside tolerance
50
B
Device not ready
(CAN)
Device or device interface not ready
51
A
No BUS-Signal from
AEC
No stop signal from automatic exposure
control
52
A
FLXIS not ready No communication to TV chain
53
B
Anode heat content
>100% !
Max Tube heat storage capacity, cool
down the tube
61
B
Receiver overflow Serial interface receive buffer overflow
62
B
Transmitter overflow Serial interface transmit buffer overflow
200 | DR 400 | X-Ray Generator Mini Console
3231E EN 20210708 1530