Minimum room height 245 cm
Tabletop attenuation equivalent mm
Aluminum
≤ 0.7
According to
DIN EN 60601-1-3 with 100kV and
HVL 3.6 mm Al
FDA 21 CFR § 1020.30 (n) with
100kV and HVL 3.6 mm Al
Weight
Fixed height radiographic table 290 kg
Elevating radiographic table 350 kg
X-ray tube stand column 120 kg
X-ray tube stand arm 25 kg
X-ray tube plus collimator (maxi-
mum weight)
40 kg
Maximum load on the radiographic
table
320 kg
Movement ranges
Transversal axis or y-axis move-
ment (back & front)
± 7 cm
Vertical axis or z-axis movement
(up & down)
33.5 cm to 180 cm from the floor
The movement range may vary depend-
ing on the type of X-ray tube.
Longitudinal axis (x-axis) move-
ment (right & left)
131 cm
Alpha axis rotation (Angle of the
X-ray tube)
±110° with mechanical detents at 0°,
±45°, ±90°
Beta axis rotation (swivel of the
X-ray tube arm around the tube
stand axis)
±90° with mechanical detents at 0°,
±45°, ±90°
Bucky horizontal movement in
the table
50 cm
234 | DR 400 | Product Information
3231E EN 20210708 1530