5. Take a sequence of exposures with the following kV values. Take one
exposure per 30 seconds.
Table 2: Sequence of exposures
Time (minutes) kV Time (minutes) kV Time (minutes) kV
0.0 50 4.0 90 8.0 130
0.5 50 4.5 90 8.5 130
1.0 60 5.0 100 9.0 140
1.5 60 5.5 100 9.5 140
2.0 70 6.0 110 10.0 150
2.5 70 6.5 110 10.5 150
3.0 80 7.0 120
3.5 80 7.5 120
DR 400 | Introduction to DR 400 | 63
3231E EN 20210708 1530