Agilent 1200 Infinity Series DAD User Manual 175
Maintenance
9
Replacing the Max-Light Cartridge Cell
5
The lever should be in the final down position. 6 Pull the cell cartridge holder completely out towards the
front.
7 Remove the cell from the cartridge holder. 8 Replace the black hoods [A] to the cell interfaces (in/out)
and insert plugs [B] for save storage.
NOTE
The hoods and the plugs should be always in place to
protect the flow cell.
For longer storage, the flow cell should be flushed and
filled with isopropanol to prevent the grow of algae.
Store it in the plastic case provided with the Max-Light
Cartridge Flow Cell.