350 Flame Photometric Detector (FPD)
Diagnostics
34 of 36
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
6. The sulfur mode can be particularly hard to light. Try changing to the
phosphorus mode flows, lighting the flame, and gradually alter the flows
to the sulfur values.