Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet 210
Leak testingEPC and manual inlets
Performing the leak testEPC inlets
Note Be sure to complete all of the preparation steps on pages 20 and 21 before
continuing.
1. Set the inlet to Split Mode.
2. Configure the column as 0 length.
Press
[Mode/Type]
Press
[Config] [Column 1] or [Config] [Column 2] and enter “0” in the first
column of the “Dim” field.