240 Programmed Temperature Vaporization Inlet (PTV)
Splitless modes
8 of 46
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Pulsed splitless mode
The pressure pulse modes (using the septum- or septumless head) increase
inlet pressure just before the beginning of a run and return it to the normal
value after a specified amount of time. The pressure pulse sweeps the sample
out of the inlet and into the column faster, reducing the chance for sample
decomposition in the inlet.