Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms 610
NPD checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
Zero 10 Pk wd 0.04
Att 2^ 7 Thresh 7
Cht sp 1 [Stop time 12.0]
Ar Rej 1000
SS/COC/PTV/PPIP
Sensitivity Azobenzene area counts 306,000
Malathion area counts 575,000
3396B or 3396C or 3397A integrator sample conditions with INET input
[6890 INET output, Range 0]
>>Set DATE and TIME<<
Zero 10 Pk wd 0.04
Att 2^ 7 Thresh 7
Cht sp 1 [Stop time 12.0]
Ar Rej 1000
SS/COC/PTV/PPIP
Sensitivity Azobenzene area counts 12,000
Malathion area counts 22,500
Agilent ChemStation
SS/COC/PTV/PPIP
Sensitivity Azobenzene area counts 38
Malathion area counts 71
Agilent Chemstation—ASTM noise
• Measured with 6890 isothermal @ 100°C
• 10 minute blank run, noise range(s) >1 minute
• Performance + noise report
NPD ASTM noise ≤ .0765 pA