Inlets
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed Inlet 220
Leak testingEPC and manual inlets
Potential leak points
Check the following areas when checking an inlet system for leaks.
In the oven
Make sure the bottom of the inlet is correctly capped.
On the inlet
Septum
O-ring under inlet septum fitting at top of inlet
Liner ferrule seal at bottom of inlet
At EPC module
Three O-rings behind block where inlet pneumatic lines enter module
Two O-rings for each valve
Septum purge cap