EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6890 Series

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #311 background imageLoading...
Page #311 background image
340 Electron Capture Detector (ECD)
Maintaining an ECD detector
24 of 24
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
6. Set the oven temperature to 250°C.
7. Allow thermal cleaning to continue for several hours, and then cool the
system to normal operating temperatures.
Performing a radioactivity leak test (wipe test)
ECDs must be tested for radioactive leakage at least every six months. Records
of tests and results must be maintained for possible inspection by the Nuclear
Regulatory Commission and/or responsible state agency. More frequent tests
may be conducted when necessary.
The procedure used is the wipe test. A Wipe Test Kit (part no. 18713-60050)
is supplied with each new ECD. Refer to the information card supplied in the
Wipe Test Kit for instructions on performing the wipe test.

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals