Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms 610
Microcell ECD checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
6890 Analog output
Noise ([N1])
Measured at [6890 Range 0],
[3396 Attn. 3, Chart speed 1, zero 50] <37 mm for 1 min. measurement
Signal
6890 Range 6
3396
Zero 10 Pk wd 0.04
Att 2^ 5 Thresh 5
Cht sp 1 Stop time 11.0
Ar Rej 100000
Area, Lindane peak 47,950 × [N1]
Agilent ChemStation
Noise ([N3])
Measured by ChemStation, ASTM [N3] < 3 Hz
Signal
Area, Lindane peak 7500 × [N3]