EasyManuals Logo

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
210 Split/Splitless Inlet
Replacement procedures
18 of 30
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
5. Remove the plumbing block from the front of the manifold (one captured
Torx T-10 screw). Replace the three rubber O-rings behind the block if
they are worn or damaged.
6. Remove the long screw (Torx T-20) from the top of the manifold and slide
the manifold out of the back of the GC.
7. Reinstallation is the reverse of removal.
Replacing the supply fitting on a Type 1 flow manifold
Carrier gas enters an inlet EPC flow manifold through a fritted, stainless steel
fitting. If the frit in this fitting clogs and cannot be cleaned, replace the
assembly as a whole. Remove the two screws attaching the fitting to the
manifold, replace the O-ring behind the fitting, and install a new fitting.
To clean the supply fitting frit, use solvents and/or an ultrasonic bath, then,
dry the supply fitting with an N
2
or air stream.
Figure 210-12 Replacing the supply fitting

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals