520 Gases
Gas requirements
2 of 14
Jun 2001
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
Gases for capillary columns
When used with capillary columns, GC detectors require a separate makeup
gas for optimum sensitivity. For each detector and carrier gas, there is a
preferred choice for makeup gas. Table 520-2 lists gas recommendations for
capillary columns.
Table 520-2 Gas Recommendations for Capillary Columns
Detector Carrier
gas
Preferred
makeup gas
Second choice Detector anode
purge or reference
gas
Electron Capture Hydrogen Argon/Methane Nitrogen Anode purge must
be same as makeup
Helium Argon/Methane Nitrogen
Nitrogen Nitrogen Argon/Methane
Argon/
Methane
Argon/Methane Nitrogen
Flame Ionization Hydrogen Nitrogen Helium Hydrogen and
air for detector
Helium Nitrogen Helium
Nitrogen Nitrogen Helium
Flame
Photometric
Hydrogen Nitrogen Hydrogen and air for
detector
Helium Nitrogen
Nitrogen Nitrogen
Argon Nitrogen
Nitrogen-
Phosphorus
Helium Nitrogen Helium** Hydrogen and
air for detector
Nitrogen Nitrogen Helium**
Thermal
Conductivity
Hydrogen* Must be same
as carrier and
reference gas
Must be same
as carrier and
reference gas
Reference must be
same as carrier and
makeup
Helium
Nitrogen

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals