EasyManua.ls Logo

Agilent Technologies 6890 Series - Dual Wavelength FPD; Operating Conditions

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 of 36
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Photometric Detector (FPD) 350
Theory of operation
Dual wavelength FPD
The dual wavelength FPD is a single burner module with two PMT housings,
one with a sulfur filter and the other with a phosphorus filter. Because the
optimum gas flows for these elements are quite different, performance of this
detector is necessarily a compromise.
Two signal channels and two electrometer boards are used, one for each PMT.
The Back Det control table runs the detector, while the Front Det operates in
a special signal only mode.
Operating conditions
In order for the FPD to operate properly, the temperature must be set above
120°C, and the air and hydrogen flows must be On.

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Related product manuals