EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6890 Series

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
230 Cool On-Column Inlet
Leak testing EPC and manual inlets
12 of 20
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Performing the leak testEPC inlets
Note Be sure that the preparation steps on pages 10 and 11 are complete before
proceeding with this test.
1. Set the inlet pressure to 25 psi on the keypad.
2. Wait approximately 15 seconds for equilibration.
If pressure cannot be achieved, either a very large leak is present in
the system, or the supply pressure is not high enough.
3. Turn the inlet pressure Off.
4. Note the Actual reading on the display and monitor the pressure for
10 minutes.
Press [Front Inlet] or [Back Inlet]. Move to the Pressure field and enter 25.
Press [Front Inlet] or [Back Inlet], scroll to the "Pressure"
field, and press [
OFF]

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals