Potential leak points
Check the following areas when checking an inlet system for leaks.
In the oven
Make sure the bottom of the inlet is correctly capped.
On the inlet
• Septum
• O-ring in top of inlet
• Lower inlet seal at bottom of inlet
At EPC module
• Three O-rings behind block where the inlet’s pneumatic lines enter
module
• Two O-rings for each valve
• Septum purge cap