EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6890 Series

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #460 background imageLoading...
Page #460 background image
1 of 6
Jun 2001
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
510 Environmental
Temperature and humidity ranges
Operating the GC within the recommended ranges insures optimum
instrument performance and lifetime.
After exposing the GC to extremes of temperature or humidity, allow 15
minutes for it to return to the recommended ranges.
Ventilation requirements
The GC is cooled by convection: air enters vents in the side panels and
underneath the instrument. Warmed air exits through slots in the top, rear,
and side panels. Do not obstruct air flow around the instrument.
Caution For proper cooling and general safety, always operate the instrument with
cover panels properly installed.
Venting oven exhaust
Hot air (up to 450°C) from the oven exits through a vent in the rear. Allow at
least 20 cm (10 inches) clearance behind the instrument to dissipate this air.
Recommended temperature range Temperature range
20 to 27°C 5 to 40° C
Recommended humidity range Humidity range
50 to 60% Up to 31°C, 5 to 80%
At 40°C, 5 to 50%
Recommended altitude range
Up to 2000 m

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals