210 Split/Splitless Inlet
Leak testing—EPC and manual inlets
20 of 30
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Figure 210-14 Capping the bottom of the inlet and septum purge vent
Note Ma ke su re t ha t t he c ar rie r ga s sou rc e pr es su re is at le ast 35 ps i. C ar ri er s ou rce
pressure should always be at least 10 psi greater than the desired inlet
pressure.
Manual
EPC
Split vent
Septum purge vent
Septum purge vent