Entering a Method
User Information 101
Setting the Last Vial
Last Vial specifies the location of the last vial to be analyzed by
this method.
Use the cursor keys to highlight Last Vial. Key in the desired
location, followed by Enter.
The allowed range for this setting is 1-70. The number of the
Last Vial cannot be lower than the number of the First Vial. If
the Last Vial location is lower than the First Vial location, the
Headspace Sampler gives an error message and the Last Vial
location must be re-entered.
Setting shake
Shake specifies the agitation level of the oven carousel during
vial equilibration. There are three shake settings: off, low, and
high. Specify the level desired using the cursor keys followed by
Enter.