Troubleshooting and Error Messages
User Information 181
Communication Between the PC and Headspace Sampler
This type of problem results from a break in the communication
between the computer and the Headspace Sampler. Check the
cable connections between the computer and the sampler.
Check the following for Headspace Samplers set up with a LAN
connection:
The Headspace Sampler and computer have compatible
network addresses. For example, the IP address are
compatible on the same network.
The Enable RS-232 setting is OFF. See Advanced Function 16
on page 83.
The LAN address is correct. See Advanced Function 14 on
page 82. Reset or power cycle the Headspace Sampler for IP
address changes to take effect.
Use the ping command from a PC command window to test
the connection.
Check the following for a Headspace Sampler with an RS-232
connection:
The com port settings are correct. Bits per second: 9600;
Data bits: 8; Parity: None; Stop bits: 1; Flow control:
Hardware.
The Enable RS-232 setting is ON. See Advanced Function 16
on page 83.
medium_standard.book Page 181 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals