The Keypad
User Information 57
Vial Parameters
Description
Select Vial Parameters from the Active Method menu to
specify the vial(s) analyzed during a method and access the
shake function.
First Vial Designates the first vial (1–70) analyzed by the
current method.
Last Vial Designates the last vial (1–70) analyzed by the
current method. The last vial number cannot be lower than the
first vial number. The Headspace Sampler will give an error
message and you will need to re-enter the vial number.
Shake Shake specifies the agitation level of the oven carousel
during vial equilibration. Shaking the vial may shorten the
needed vial equilibration time.
There are three shake settings: off, low, and high. Specify the
level desired using the cursor keys followed by Enter. Low is
sufficient for most applications.
First Vial 1
Last Vial 22
SET
VIAL PARAM.
Shake Off
Not Ready