Troubleshooting and Error Messages
User Information 175
240 Sample probe lifter error
This error results when the sampler lifter motor encoder senses
an obstruction or cannot find its home position.
Check and clear any obstruction to the lifter.
Make sure vial size setting is correct. See Advanced
Function 12 on page 80.
Check the carousel alignment. See page 142.
See Recovering from a vial delivery error on page 177. Call an
Agilent service representative with the number and name of the
error.
250 Carousel error
This error results when the carousel motor encoder senses an
obstruction or cannot find its position.
Check and clear any obstruction to the carousel.
Check the carousel alignment. See page 142.
See Recovering from a vial delivery error on page 177. Call an
Agilent service representative with the number and name of the
error.
260 6-port valve error
This error results when the 6-port valve motor encoder senses
an obstruction or cannot find its home position.
Check and clear any obstruction to the valve motor linkage.
Check the for any rubbing on the back side of the insulated
cover to the loop heated zone.
See Recovering from a vial delivery error on page 177. Call an
Agilent service representative with the number and name of the
error.
medium_standard.book Page 175 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals