Troubleshooting and Error Messages
User Information 185
Power ON, No Response
This symptom results when the line voltage from the wall outlet
does not reach the internal power supplies of the instrument.
Check the following:
Headspace Sampler line module fuses are open or blown.
Check the power selector switch setting. Fuses will blow if
switch is set to 115 VAC and the line voltage is 220-260 VAC.
Power cord not connected.
Power ON switch, set to OFF.
Line voltage not available at electrical outlet.