84 User Information
Advanced Functions
Vent Valve purging
Description
V2 Sequence Purge This function sets the amount of time that
the V2 vent valve is open after each vial injection. The default
value is 30 second.
V2 Standby Purge This function opens the vent valve and
allows pressurization gas to escape through the vent. Use this
function to remove any traces of unwanted material from the
Headspace Sampler plumbing before running a method, if
desired.
V2 Standby Purge
Vent Purge
V2 Sequence Purge
Not Ready
ACT
Start purging
PURGE VALVES
Timer (min)
RUNNING
0.0 1
SET